* | The recommended holding time for this analysis is only 15 minutes. The sample was analyzed as soon as it was possible but it was received and analyzed past holding time. |
** | The recommended holding time for field filtering is only 15 minutes. The sample was filtered as soon as possible but it was filtered past holding time. However, the sample was analyzed within holding time. |
_NONE | No Qualifiers |
_SeeAtt | See Attachment |
A-01 | [Custom Value] |
AN-IP | Sample results for structural isomers may have contribution from their isomeric pair. |
B | Blank contamination. The analyte was found in the associated blank as well as in the sample. |
B-01 | This analyte was found in the method blank, which was possibly contaminated in the lab during preparation. The reporting limit was raised due to the contamination. |
B-02 | This analyte is detected in the method blank below the MRL, but above the method acceptance criteria. |
B-03 | Dilution water blank exceeds 0.20mg/L. |
B-04 | This analyte was found in the travel blank, which was possibly contaminated during transportation and/or preparation. The batch was accepted since this analyte was not detected for all the samples in the batch. |
B-06 | This analyte was found in the method blank, which was possibly contaminated during sample preparation. The batch was accepted since this analyte was either not detected or more than 10 times of the blank value for all the samples in the batch. |
B-08 | Analyte is found in the method blank, which was possibly contaminated during sample preparation. |
B-field | No field blank was either received or specified in this batch. Therefore, samples were analyzed without field blank. |
BOD-01 | The sample dilutions set-up for the BOD analysis did not meet the oxygen depletion criterion of at least 2 mg/l, therefore the reported result is an estimated value only. |
BOD-02 | The sample dilutions set up for the BOD analysis did not meet the criterion of a residual dissolved oxygen of at least 1 mg/l, therefore the reported result is an estimated value only. |
BOD-03 | The sample dilutions set-up for the BOD analysis did not meet the final DO reading with a value of equal or greater than 1 mg/l, therefore the reported result is an estimated value only. |
BOD-04 | The sample dilutions set-up for the BOD analysis did not meet the final DO reading with a value of equal or greater than 1 mg/l, due to matrix interferences.. |
BR | Analyte was found in the method blank, which was possibly contaminated in the lab during preparation. The reporting limit was raised to account for the contamination. |
BS-01 | The recovery of this analyte in the BS/LCS was over the control limit due to a possible contamination. The batch was accepted based on another acceptable BS and/or MS and MSD that meet the BS criteria. |
BS-02 | The recovery of this analyte in the BS/LCS was outside the control limits due to cross contamination from highly contaminated samples during preparation. |
BS-03 | The recovery of this analyte in the BS/LCS was outside the control limits. The sample result was accepted based on another acceptable BS/LCS and/or MS and MSD that meet BS criteria. |
BS-04 | The recovery of this analyte in LCS or LCSD was outside control limit. Sample was accepted based on the remaining LCS, LCSD or LCS-LL. |
BS-H | The recovery of this analyte in the BS/LCS was over the control limit. Sample result is suspect. |
BS-L | The recovery of this analyte in the BS/LCS was below the control limit. Sample result is suspect. |
C-3620 | The "Florisil Cleanup" was performed to the sample. |
C-3640 | The "GPC Cleanup" was performed to the sample. |
C-3665 | The "Sulfuric Acid Cleanup" was performed to the sample. |
CHMT-01 | The sample was analyzed directly with acid dilution. |
CN-1 | See case narrative for an explanation of results. |
DI_WET | On Deionized Water W.E.T extract (STLC). |
DryW_Clnt | The dry weight result was given by client. |
E | The concentration indicated for this analyte is an estimated value above the calibration range of the instrument. This value is considered an estimate (CLP E-flag). |
E-01 | The concentration indicated for this analyte is an estimated value above the calibration range. |
F-01 | The sample was received without proper field flocculation. Therefore, flocculation process was performed at the lab as soon as possible. |
FILT | The sample was filtered prior to analysis. |
FP | Formation Potential |
HC-02 | Hydrocarbon pattern present in the requested fuel quantitation range but does not resemble the pattern of the requested fuel. |
I | Internal Standard outside of acceptance limit due to possible matrix effects |
I-03 | Low internal standard recovery possibly due to matrix interference or leak in system. The result is suspect. |
I-05 | Low internal standard recovery possibly due to matrix interference. The result is suspect. |
J | Estimated conc. detected MDL. |
M | Sample result is matrix suspect. |
M-02 | Due to the nature of matrix interferences, sample was diluted prior to preparation. The MDL and MRL were raised due to the dilution. |
M-03 | Due to insufficient sample volume, sample was diluted prior to preparation. The MDL and MRL were raised due to the dilution. |
M-04 | Due to the nature of matrix interferences, sample extract was diluted prior to analysis. The MDL and MRL were raised due to the dilution. |
M-05 | Due to the nature of matrix interferences, sample was diluted prior to analysis. The MDL and MRL were raised due to the dilution. |
M-06 | Due to the high concentration of analyte inherent in the sample, sample was diluted prior to preparation and/or analysis. The MDL and MRL were raised due to this dilution. |
M-08 | Incomplete sample loading during extraction due to sample matrix |
MI | Manual Integration was performed |
MI-01 | Manual intergration is necessary due to irregular baseline. |
MI-02 | Manual intergration is necessary due to low signal to noise ratio. |
MIC-2 | Result is suspect due to QC failure. |
MS-01 | The spike recovery for this QC sample is outside of established control limits possibly due to sample matrix interference. |
MS-02 | The RPD and/or percent recovery for this QC spike sample cannot be accurately calculated due to the high concentration of analyte inherent in the sample. |
MS-03 | Multiple analyses indicate the percent recovery is out of acceptance limits due to a possible matrix effect. |
MS-04 | Visual evaluation of the sample indicates the RPD or QC spike is above the control limit due to a non-homogeneous sample matrix. |
MS-05 | The spike recovery and/or RPD were outside acceptance limits for the MS and/or MSD due to possible matrix interference. The LCS and/or LCSD were within acceptance limits showing that the laboratory is in control and the data is acceptable. |
MS-06 | Due to noted non-homogeneity of the QC sample matrix, the MS/MSD did not provide reliable results for accuracy and precision. Sample results for the QC batch were accepted based on LCS/LCSD percent recoveries and RPD values. |
MS-07 | The spike recovery was outside acceptance limits for the MS and/or MSD. The batch was accepted based on acceptable LCS recovery. |
MS-08 | Due to the nature of matrix interferences, sample was diluted prior to analysis. The MS/MSD could not be quantitated due to the dilution. The batch was accepted based on acceptable LCS recovery. |
MS-09 | The recoveries of MS/MSD are not valid due to high sample background |
MS-10 | Due to insufficient sample, LCS/LCSD were analyzed in place of MS/MSD. |
MS-11 | The QC limits for MS/MSD are not applicable due to positive sample background. |
MS-BG | The spike recovery was outside of QC acceptance limits for the MS and/or MSD due to sample background. The QC batch was accepted based on LCS and/or LCSD recoveries within the acceptance limits. |
O-02 | Sulfides present in the sample. Therefore, 24 hours holding time for the analysis already expired. |
O-04 | This analysis was performed outside the EPA recommended holding time. |
O-05 | The extraction for this analyte was performed outside of the EPA recommended holding time. |
O-07 | Sample date and/or time not provided by client. Therefore, default date and/or time has been entered. The analysis may be outside of recommended holding time. |
O-08 | The original extraction and/or analysis of this sample yielded QC recoveries outside acceptance criteria. It was re-extracted/re-analyzed after the recommended maximum hold time. |
O-09 | This sample was received with the EPA recommended holding time expired. |
O-10 | The original analysis of this sample yielded QC recoveries outside acceptance criteria. It was re-analyzed after the recommended maximum hold time. |
O-11 | The sample was originally analyzed within holding time. However, it required a dilution and the re-analysis was performed after the recommended holding time had expired. |
O-12 | The sample was originally analyzed within holding time. However, it was reanalyzed without dilution that exceeded the recommended holding time. |
O-13 | The sample was originally analyzed within hold time. However, the re-analysis was performed due to possible carry over and/or cross contamination after the recommended holding time had expired. |
O-14 | This analysis was requested by the client after the holding time was exceeded. |
O-15 | The sample was received with the recommended holding time nearly expired. It was analyzed as soon as possible but the maximum holding time was slightly exceeded. |
O-20 | As per vial label, this sample was received with HCl preservation, however sample pH was found to be >2 after VOC analysis possibly due to matrix effect or loss of acid during sampling. |
O-21 | This sample was analyzed 1 hour past the EPA recommended holding time. |
O-22 | This sample was analyzed 2 hours past the EPA recommended holding time. |
O-25 | This sample was received unpreserved and with the recommended holding time for preservation of 48 hours expired. |
O-26 | This sample was received unpreserved and with the recommended holding time for filtration of 24 hours expired. |
O-27 | The sample was received in the improper sampling container. |
P-01 | Low recovery due to preservative. Sample data accepted based on passing LCS result. |
P-2 | Sample received without proper preservation and was preserved at the lab upon receiving. |
P-3 | The sample was preserved with ascorbic acid, but the pH was >2 possibly due to no, or insufficient preservation with HCl. The sample was not analyzed within 24 hours, as required by method for sample with pH>2. |
P-6 | The sample was filtered and preserved prior to analysis. |
P-7 | The sample was preserved within 14 days. |
Q | One or more quality control criteria failed. |
Q-01 | The recovery of this analyte in QC sample was outside control limits. Sample was justified as ND based on the low level standard at or below the reporting limit. |
Q-02 | Low recovery of this analyte in the QC sample. The analysis of the low level standard produced acceptable recovery indicating that the sample result might be accurately reported as Not Detected. |
Q-08 | High bias in the QC sample does not affect sample result since analyte was not detected or below the reporting limit. |
Q-09 | This analyte bias high in QC sample. A fresh spiking solution is going to be prepared. |
Q-10 | This analyte is high bias in QC samples, sample result is suspect. |
Q-11 | This analyte is low bias in QC samples, sample result is suspect. |
Q-12 | The RPD result exceeded the QC control limits; however, both percent recoveries were acceptable. Sample results for the QC batch were accepted based on the percent recoveries and/or other acceptable QC data. |
QC-2 | This QC sample was reanalyzed to complement samples that require re-analysis on different date. See analysis date. |
Q-H-1 | High bias, data was accepted since sample was not detected. |
Q-L-03 | This analyte is low in QC sample. Sample data is accepted based on acceptable CCVs. |
Q-ME | Acceptable QC with marginal exceedance |
Q-R-01 | Analyses are not controlled on RPD values from sample concentrations less than the reporting limit. QC batch accepted based on LCS and/or LCSD QC results. |
QR-03 | The RPD value for the sample duplicate or MS/MSD was outside of QC acceptance limits due to matrix interference. QC batch accepted based on LCS and/or LCSD recovery and/or RPD values. |
QR-04 | The RPD value for the MS/MSD was outside of QC acceptance limits however both recoveries were acceptable. The QC batch was accepted based on acceptable results for the recoveries and RPD for the LCS and LCSD. |
QR-BS | The RPD value for the BS/BSD (LCS/LCSD) was outside of QC acceptance limits however both recoveries were acceptable. The QC batch was accepted based on acceptable results for the recoveries of the BS (LCS) and BSD (LCSD). |
Q-S | QC samples were not performed due to insufficient sample received. |
R-01 | The MDL and/or MRL for this analyte has been raised to account for matrix interference. |
R-02 | The RPD was outside of QC acceptance limits due to possible matrix interference. |
R-03 | The RPD is not applicable for result below the reporting limit (either ND or J value). |
R-04 | Due to foaming, the sample was diluted prior to analysis. The reporting limits were raised due to the dilution. |
R-07 | Incomplete sample loading during extraction due to sample matrix. The MDL and MRL were adjusted accordingly. |
RD | Data was recalled. See revised report for corrected results. |
RE1_P | % Residual error at lowest calibration point fails over 50% allawable limit. This only affects sample concentration at or near this point. However, it will not have a significant affect beyond this point. |
R-MS | Results reported using MS/MS as the primary detector. |
S-01 | The surrogate recovery could not be calculated due to sample dilution required from high analyte concentration and/or matrix interferences. |
S-02 | The surrogate recovery for this sample cannot be accurately quantified due to interference from coeluting organic compounds present in the sample extract. |
S-03 | High surrogate recovery for this sample is possibly due to a sample matrix effect. The data was accepted since all target analytes were not detected. |
S-04 | The surrogate recovery for this sample is outside of established control limits due to possible sample matrix effect. |
S-05 | Surrogate recovery was below acceptance limit possibly due to matrix effect. Sample data was justified as acceptable since all target analytes were still not-detected or below the reporting limits when adjusted accordingly to surrogate recovery. |
S-06 | The recovery of this surrogate is outside control limits due to sample dilution required from high analyte concentration and/or matrix interference's. |
S-07 | Surrogate recovery out of acceptance limits for this sample is possibly due to sample matrix effect, confirmed by re-extracting and/or re-analyzing the sample. |
S-08 | No surrogate recovery, possibly surrogate spiking was missed. |
S-09 | Wrong amount spiked, quantification is not accurate |
S-10 | Surrogate recovery outside method QC limits due to extraction related problems |
S-11 | Surrogate recovery outside of control limits. The data was accepted based on valid recovery of the remaining surrogate. |
S-AC | Acid surrogate recovery outside of control limits due to a possible matrix effect. The data was accepted based on valid recovery of remaining two acid surrogates. |
S-BLK | Surrogate recovery outside of control limits for Method Blank. The data was accepted since all target analytes were not detected |
S-BN | Base/Neutral surrogate recovery outside of control limits due to a possible matrix effect. The data was accepted based on valid recovery of remaining two base/neutral surrogates. |
S-BS | Surrogate recovery outside of control limits for LCS. The data was accepted based on valid recovery of the target analytes. |
S-GC | Surrogate recovery outside of control limits due to a possible matrix effect. The data was accepted based on valid recovery of the remaining surrogate. |
S-HI | High surrogate recovery was confirmed as a matrix effect by a second analysis. |
S-LOW | Low surrogate recovery confirmed as a matrix effect by a second analysis. |
S-MS | Surrogate recovery outside of control limits for MS/MSD. The data was accepted based on valid recovery of the target analytes. |
S-MS1 | Surrogate recovery outside of acceptance window confirmed as matrix effect by analysis of MS/MSD on this sample. |
Supp | This result has been revised from the original report. |
T-AgBaH | The sample was treated with Silver, Barium and H+ cartridges to minimize chloride and sulfates interferences prior to analysis. |
TOX-01 | Due to sample matrix interferences, second column measurement exceeding 10% of the two-column total. |
T-RP | The sample was treated with Organics cartridges to minimize organic interferences prior to analysis. |
U-01 | The sample was received without the proper preservation. |
U-02 | The sample was received at the lab without proper preservation. However, the sample was then preserved at the lab. |
U-03 | Due to limited sample volume provided, the method criteria cannot be achieved. |
VTR-01 | The sample was filtered with 1.5 micron filter at lab. |
WC1 | COC was not provided upon receipt |
WC2 | COC was not properly completed |
WC3 | COC and sample labels do not match |
WP2 | pH out of acceptable range |
WP2A | pH Lot, pH Value |
WP2B | Acid Lot, Acid Amount (pH Adjustment) |
WP3 | Free Chlorine >= 0.1 |
WP3A | Cl Lot |
WS1 | Sample(s) were not received on ice as required |
WS2 | Sample(s) were not intact |
WS3 | Sample(s) were not using their proper container |
WS4 | Sample(s) does not have sufficient sample volume |
WS4V | Headspace was present for VOC zero headspace methods |
WS5 | Sample(s) were not received within hold time |